RM-2030 portable X - gamma radiometer for measuring irradiation dose and dose equivalent rate, absorbed dose rate of intelligent instrumentation, in addition to be able to measure the high energy, low energy gamma rays, but also for low energy X ray were accurate measurements. Widely used in environmental monitoring, health and epidemic prevention, import and export commodity inspection, medical radiation, building materials, petroleum chemical, geological survey, scrap iron and steel, nuclear laboratories in the field of radiation protection monitoring, and airport, railway stations, ports of radioactive security.
characteristic:
1, in line with national standards of energy response and solve the previous radiometer for energy response indicators unqualified caused by the same activity (due to the measured energy different) the difference between the results of ten times or even several times the error of measurement.
2, compared with foreign imports of similar instruments, with a higher performance price ratio.
3, the use of NaI crystal, high sensitivity.
4, graphic liquid crystal display, integrated intelligent design, easy to operate.
5, flexible choice of measuring time.
6, according to the working situation set alarm threshold.
7, measurement units can be switched between uSv/h, uR/h, uGy/h.
Technical index:
1, energy response: 48Kev ~ 3MKev less than 30% (normalized to 137Cs)
2, energy threshold: 40Kev.
3, measurement range: 0.01uSv/h ~ 200uSv/h.
4, measurement accuracy: to set the reliability of 95%, a reading:
0.50 ~ 50uSv/h is less than 5%
50 ~ 200uSv/h is less than 10%
5, angular response: the instrument in the 90o ~ 0o does not exceed 25% of the response in the calibration direction.
6, the reading display: five digits to uSv/h, uR/h, uGy/h as the unit gives the measurement results.
7, alarm threshold: available in 0.25uSv/h, 2.5uSv/h, 25uSv/h arbitrary choice.
8, measuring time:
(1) a second measurement, display the current average value of three seconds
(5) each measurement of 2 seconds to 100 seconds, showing the measured time per second average.
9, the use of the environment: temperature -10 ~ +50 DEG C, relative humidity: less than or equal to 98% (40 C)
10, under voltage indication: when the battery voltage drops to 2V, indicating the replacement of the battery.
11, power supply voltage: two 1.5V 1 battery.
12, power consumption: less than 200mW.
13, weight: about 1Kg (including two battery weight).
14, size: length * width * high =425mm * 80mm * 210mm.